Test Items:
|
22 parameters: RBC, HGB, WBC, LY#,
MO#, GR#, LY%, MO%, GR%, HCT, MCV, MCH, MCHC, RDW-CV, RDW-SD, PLT, MPV, PCT,
PDW |
|
Histogram Display: |
White Blood Cell Histogram
(Colour) |
Red Blood Cell Histogram (Colour) |
Platelet Histogram (Colour) |
Linearity Range:
|
WBC (×109/L): 0.0-99.9 |
RBC (×1012/L): 0.00-99.9 |
HGB (g/L): 20-200 |
MCV (fL): 40.0-150 |
PLT(×109/L): 0.0-999 |
Precision (CV):
|
WBC (×109/L) < 2%
(5.0-10.0×109/L) |
RBC (×1012/L) <1.5%
(3.5-6.0×1012/L) |
HGB (g/L) <1.5%
(110-180g/L) |
MCV (fL) <0.5%
(80.0-110fL) |
PLT (×109/L) <4.0%
(100-500×109/L) |
Data Storage:
|
25,000 results |
Dilution Ratio:
|
WBC / HGB 1:500 |
RBC / PLT 1:50000 |
Sample volume:
|
Periphery/whole blood: 20µl |
Display:
|
640×480 pixels, 9.4" TFT |
Printer:
|
Built-in thermal printer, paper width:
5.7cm |
Test Time:
|
Less than 40 seconds |
Continuous Work Time:
|
More than 8 hours |
Power:
|
220V±22V 50Hz±1Hz |
Fuse:
|
BLX-1 50×20-2A |
Input Power:
|
≤180VA |
Interfaces:
|
RS232
joint |
|
|
|
|
|
|